Assembly Replacement
A38 Lowband Switch Filter
A38 Lowband Switch Filter
Tools Required
— none
Removal Procedure
Refer to
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
3. Position the signal generator with the front panel facing you.
4. Lift the retention levers and disconnect the board assembly from the A31
5. Lift the board assembly out of the slot.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post–repair adjustments and performance tests that pertain
E8257D/67D, E8663D PSG Signal Generators Service Guide
Figure 3-71
for this procedure.
Cover" on page 258
and
Motherboard.
to this removal procedure.
"Inner Instrument Cover" on page
"Outer Instrument
260.
377
Need help?
Do you have a question about the E8257D and is the answer not in the manual?