Test 2-5: AC Flatness - 250 kHz Filter
Test 2-5: AC Flatness - 250 kHz Filter
Description
The purpose of this test is to verify that the AFG meets its specifications for AC
flatness with the 250 kHz filter enabled.
Equipment Setup
Figure 2-6 Equipment Setup for Test 2-5 and Test 2-6
Test Procedure
1 Reset the AFG:
2 Set up the AFG to output a 24 dBm sine wave with the 250 kHz filter enabled:
Keysight E1445A Arbitrary Function Generator Service Guide
– Connect equipment as shown in Figure 2-6
– Set DMM to ACV, autorange
*RST;*CLS
Reset AFG and clear status registers
Verification Tests
51