Assembly Replacement
A3 Power Switch
A3 Power Switch
Tools Required
— T–10 driver
Removal Procedure
Refer to
< US/MY5XXX3XXX.
Refer to
≥ US/MY5XXX3XXX.
1. Disconnect the power cord.
2. Remove the front panel. Refer to
3. Disconnect the W8 ribbon cable from the A3 Power Switch by lifting the
4. Using the T–10 driver, remove the four screws (12) from the A3 Power
5. Remove the A3 Power Switch.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque all T–10 screws to 9 in–lbs.
3. Perform the post–repair adjustments and performance tests that pertain