A35 3-20 Ghz I/Q Modulator (E8267D Only - Standard Or With Option 015); Tools Required; Removal Procedure; Replacement Procedure - Keysight Technologies E8257D Service Manual

Psg signal generators
Hide thumbs Also See for E8257D:
Table of Contents

Advertisement

Assembly Replacement
A35 3–20 GHz I/Q Modulator (E8267D Only - Standard or with Option 015)
A35 3–20 GHz I/Q Modulator (E8267D Only - Standard or with
Option 015)

Tools Required

— T–10 driver
— 5/16" open–ended wrench
— needle–nose pliers

Removal Procedure

Refer to
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
3. Position the signal generator with the A31 Motherboard facing you and the
4. Using the 5/16" open–ended wrench, disconnect the following RF cables:
5. Using the needle–nose pliers, disconnect the following flexible cables:
6. Using the T–10 driver, remove the three screws (1) that attach the A35 I/Q
7. Disconnect the W70 ribbon cable from the A35 I/Q MOD.
8. Remove the A35 I/Q MOD from the RF deck.

Replacement Procedure

— Reverse the order of the removal procedure. (See the internal cover overlay
E8257D/67D, E8663D PSG Signal Generators Service Guide
Figure 3-61
for this procedure.
Cover" on page 258
and
RF deck on top.
— W105 from A35J4
— W77 from A35J1
— W78 from A35 I/Q MOD
— W104 from A35J3
— W62 from A35J7
— W63 from A35J6
— W79 from A35J9
— W80 from A35J106
— W97 from A35J8
MOD to the RF deck.
for cable color codes.)
"Inner Instrument Cover" on page
"Outer Instrument
260.
369

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8663dE8267d

Table of Contents