Keysight Technologies E8257D Option H30 User’s Guide Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com.
Description Description The E8257D Option H30 adds a mixer and coax switch to enable the E8257D to upconvert the RF signal (input is provided on the rear panel) to a higher frequency which will be the sum and difference of the RF input signal and frequency set on the E8257D.
E8257D Option H30 Nominal Performance Characteristics Nominal Performance Characteristics Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty. Table 1 Option H30 with Option 520 (20 GHz)
Page 8
E8257D Option H30 Nominal Performance Characteristics Figure 1 Upconverted RF Flatness IF Input frequency vs. flatness over 1 GHz (dB) for fundimental mixing Figure 2 Upconverted RF Conversion Loss Conversion Loss from IF input power level (dB) for fundimental mixing 9 12 15 18 21 24 27 30 33 36 39 41 44 46 User’s Guide E8251-90072...
E8257D Option H30 Operation and Performance Verification Operation and Performance Verification Documentation Conventions The following key conventions are used throughout this document. • are labeled front panel keys. [HARDKEYS] • are unlabeled keys whose function is indicated on the instrument display.
11900B (or comparable) Step 1. Set the RF Source power level to –10 dBm and frequency to 6 GHz. Step 2. Connect the RF Source to the E8257D Option H30 rear panel RF Upconverter Input, using an RF cable and adapters....
Refer to Figure 4 through Figure 7 Verify that the E8257D Option H30 Upconverter switch is working: MID Board or Coaxial Switch The following procedure will aid in determining whether the MID board or the coax switch (87222E) is functional.
E8257D Option H30 Troubleshooting Upconverter Mixer The following procedure will aid in determining if the upconverter mixer is functional: Possible damage may occur if RF upconverter input levels exceed 20 dBm or if CAUTION reverse power is applied at the RF output connector. First verify all cable configurations are correct before applying a signal to the instrument.
Declaration of Conformity A copy of the Declaration of Conformity is available upon request, or a copy is available on the Keysight Technologies web site at http://regulations.keysight.com/DoC/search.htm Statement of Compliance This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition.
E8257D Option H30 Safety and Information Before Applying Power Verify that the premises electrical supply is within the range of the instrument. The instrument has an autoranging power supply. If this prod uct is not used as specified, the protection provided by the equipment WARNING could be impaired.
E8257D Option H30 Safety and Information For continued protection against fire hazard replace line fuse only with same WARNING type and rating. The use of other fuses or material is prohibited. These servicing instructions are for use by qualified personnel only. To avoid WARNING electrical shock, do not perform any servicing unless you are qualified to do so.
E8257D Option H30 Regulatory Information Regulatory Information This section contains information that is required by various government regulatory agencies. Instrument Markings The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
Keysight for repair. If you wish to send your instrument to Keysight Technologies for service or repair: • Include a complete description of the service requested or of the failure and a description of any failed test and any error message.