ECOTEST MKS-05 TERRA Operating Manual page 55

Dosimeter-radiometer
Hide thumbs Also See for MKS-05 TERRA:
Table of Contents

Advertisement

where
is a numeric value of the measured surface beta-particles flux density in
φ
β
2
part./(cm
·min).
3.2.4.6 Presentation of verification results
3.2.4.6.1 Positive results of primary or periodic verification are registered as follows:
1) primary verification is registered in the "CERTIFICATE OF ACCEPTANCE" section;
2) periodic verification is registered in the issued Certificate of the established form.
Primary verification results are registered in Table 3.3.
3.2.4.6.2 The dosimeters that do not meet the requirements of the verification procedure
are not allowed for manufacture and use, and get the Certificate of Inadequacy.