ECOTEST MKS-05 TERRA Operating Manual page 49

Dosimeter-radiometer
Hide thumbs Also See for MKS-05 TERRA:
Table of Contents

Advertisement

Calculate main relative error of measurement in percentage.
Place the holder and the dosimeter in the position, where DER from the
8.0 µSv/h.
Perform five measurements of DER, and register the received readings in the protocol.
Calculate the DER value in µSv/h according to the formula (1).
Calculate main relative error of measurement in percentage.
Place the holder and the dosimeter in the position, where DER from the
80.0 µSv/h.
Perform five measurements of DER, and register the received readings in the protocol.
Calculate the DER value in µSv/h according to the formula (1).
Calculate main relative error of measurement in percentage.
Place the holder and the dosimeter in the position, where DER from the
2
8·10
µSv/h.
48
137
Cs source is
137
Cs source is
137
Cs source is

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents