Figure 3-20:
Connect the lead set to the yellow and white wires.
Note
If you want to communicate with the THUM adapter, ensure the Auto-Connect option for the Field
Communicator Settings is disabled. This allows the Field Communicator application to detect both the
THUM adapter and the transmitter, and let you select which one to communicate with.
A.
B.
Device connection wizard prompts
Do you intend to provide power from the Trex unit?
Leads are attached to:
Change Polling Option?
User Guide
Communicate with the THUM adapter attached to a powered HART
transmitter
Optional connection if the yellow and white wires are inaccessible.
Voltage source.
Field Communicator application
Answer
No
Transmitter
Tap Yes and select the address for the
THUM adapter, the device, or "0,63"
for both.