Intrinsically Safe Electrical Parameters - Emerson AMS Trex User Manual

Hide thumbs Also See for AMS Trex:
Table of Contents

Advertisement

Specification
Shock
B.5

Intrinsically Safe electrical parameters

Table 1: Device Communicator communication module
Ui
Ii
Pi
Ci
Li
Uo
Io
Po
Co
Lo
Table 2: Device Communicator Plus communication module
Ui
Ii
Pi
Ci
Li
Uo
User Guide
Value
The unit shall survive a 1-meter, all sides and corners drop test onto
concrete.
FOUNDATION
fieldbus
(non-FISCO)
FF + and -
30 Vdc
380 mA
1.3 W
0
0
1.89 V
32 µA
61 µW
14.3 µF
100 mH
FOUNDATION
fieldbus
mA
interface
(non-FISCO)
FF pwr
mA
and F-
FF + and -
30 Vdc
17.5 Vdc
200 mA
380 mA
380 mA
1.0 W
1.3 W
0
231 nF
0
0
0
17.31 V
FOUNDATION
(FISCO)
FF + and -
30 Vdc
215 mA (IIC) 380 mA (IIB)
1.9 W (IIC) 5.3 W (IIB)
0
0
1.89 V
32 µA
61 µW
14.3 µF
100 mH
®
HART
HART +
HART +
pwr
30 Vdc
30 Vdc
200 mA
200 mA
1.3 W
1.0 W
0
0
0
0
1.89 V
25.69 V
Technical specifications
fieldbus
HART
HART + and -
30 Vdc
200 mA
1.0 W
0
0
1.89 V
32 µA
61 µW
14.3 µF
100 mH
FOUNDATION
(FISCO)
FF pwr and
and -
F -
30 Vdc
17.5 Vdc
380 mA
215 mA (IIC)
380 mA (IIB)
1.0 W
1.3 W
0
231 nF
0
0
1.89 V
17.31 V
®
fieldbus
FF + and -
30 Vdc
1.9 W (IIC)
5.3 W (IIB)
0
0
1.89 V
187

Advertisement

Table of Contents
loading

Table of Contents