Emerson AMS Trex User Manual page 60

Hide thumbs Also See for AMS Trex:
Table of Contents

Advertisement

Field Communicator application
Power Status Message
Voltage Detected on HART
Voltage Detected on HART+
pwr
Low Voltage Detected on HART Typical status for connecting:
Low Voltage Detected on HART
+ pwr
No Voltage Detected
Voltage Detected on FF
Low Voltage Detected on FF
Voltage Provided from FF pwr
Voltage Provided from HART+
pwr
52
Description
Typical status for connecting across HART device's
communication terminals when the device is on an externally-
powered loop.
The detected voltage is greater than or equal to 8 volts.
The detected voltage is greater than or equal to 8 volts.
To a 4-wire HART device
To a WirelessHART device
Across a resistor
Across HART control system modules
The detected voltage is between 0.5 and 8 volts.
Note
You may see this status when the AC adapter is attached to the
Trex unit. If low voltage is detected, disconnect the AC adapter.
Note
This message can also appear if the Trex unit connects to a
device that was recently powered. Some devices may hold
voltage for several minutes after being disconnected from its
power supply. The Trex unit provides a warning message that it
detects voltage.
If you try to power a device when this occurs, the Trex unit asks if
you want to ignore the voltage warning and proceed with
sourcing power. If you select Yes, the Trex unit powers the
device.
The detected voltage is between 0.5 and 8 volts.
Typical status for connecting:
An unpowered device.
To some types of 4-wire HART devices.
To some types of WirelessHART devices.
This status is also displayed when the lead set is not connected to
a device, or when the Trex unit is connecting to a device.
Typical status for connecting to an externally-powered fieldbus
segment.
The detected voltage is greater than or equal to 8 volts.
The measured voltage is less than 8 volts.
The typical status when the Trex unit is powering the
FOUNDATION fieldbus device. The detected voltage is greater
than or equal to 8 volts.
The typical status when the Trex unit is powering the HART
device. The detected voltage is greater than or equal to 8 volts.