C A L I B R A T I O N
C U R R E N T OU T P U T
s e l e c t :
3
C U R R E N T O U T P U T
S E T
0 mA
s e l e c t :
C U R R E N T O U T P U T
0 mA s e t !
< e s c >
t o
C U R R E N T O U T P U T
S E T
4 mA
s e l e c t :
C U R R E N T O U T P U T
4 mA s e t !
< e s c >
t o
C U R R E N T O U T P U T
S E T
2 0 mA
s e l e c t :
C U R R E N T O U T P U T
2 0 mA s e t !
< e s c >
t o
Satellite XT 4-20 mA Versions
The 4-20 mA signal output range of the instrument is
preprogrammed to correspond to the nominal range
of the target gas with 4 mA representing the zero
o f
3
concentration and 20 mA representing the full scale
concentration.
This submenu provides the possibility to perform
functional tests and to scale external control systems.
To select a submenu actuate the <set> key to enter or
use the cursor keys <up> and <down> to move back-
wards or forwards.
1
o f
3
Allows the user to simulate a fault condition. Use the
<set> key to enter.
An analog signal output of 0 mA is now transmitted to
external control systems. Use the <esc> key to exit.
e x i t
!
Allows the user to scale external control systems with
an output signal of 4 mA representing the zero
concentration value.
2
o f
3
Use the <set> key to enter.
An analog signal output of 4 mA is now transmitted to
external control systems.
Use the <esc> key to exit.
e x i t
!
Allows the user to scale external control systems with
an output signal of 20 mA representing the full scale
concentration value.
3
o f
3
This submenu also serves to simulate alarm conditions.
Use the <set> key to enter
An analog signal output of 20 mA is now transmitted to
external control systems.
Use the <esc> key to exit.
e x i t
!
Calibration - Section 5

Current Output

Set 0 mA

0 mA set !

Set 4 mA

4 mA set !

Set 20 mA

20 mA set !
5-5