Tubing Connections; Data Link; Power Connection - Honeywell Satellite XT 9602-0200 4-20 mA Operation Manual

Satellite xt 4-20 ma series
Table of Contents

Advertisement

10.6

Tubing Connections

Tubing material is PTFE or PFA with an outer diameter of 1/4" and an inner diameter of 3/16".
Length and inner diameter of the sample line will affect the response time. Length of the
sample lines should be kept as short as possible. With a sample line ID of 3/16" (4.7 mm) and
a length of 10 m, the delay in response does not exceed 30 seconds.
Moisture may occur either as rain entering a line at an outdoor sampling location or as
condensation from outdoor-to-indoor temperature fluctuations. In case of unusual quantities of
moisture, appropriate steps must be taken to protect the instrument; consult us for respective
assistance.
Sample Line Inlet
The fitting for connecting the sample line inlet is located on top of the housing and marked
"Inlet". Improper installation of the sample inlet tube may result in dilution or even complete
loss of sample.
Sample Line Filter
Dust may be a result of construction as well as manufacturing activities. If unusual quantities
of dust are expected at the monitoring location, an appropriate filter for the sample line must
be installed. The use of a dust filter depends on the target gas, refer to the table Sensor Order
Information in Section 8, Reference Information, or contact our service department. As excess
amounts of dirt in the filters will reduce the sample flow and will thus affect concentration
readings of the instrument, make sure to regularly change sample line filters (every 1 to 6
months, depending on the ambient conditions). For order information refer to Section 8,
Reference Information, Spares and Accessories.
Module Connections
sample tubing.
Module Connections
sensor connection between Extractive
Module XT and Satellite XT.
sample tubing.
data link connecting Pyrolyzer Module
sensor connection between Extractive
XT and Extractive Module XT / Satellite
Module XT and Satellite XT.
XT.
data link connecting Pyrolyzer Module XT
10.7

Data Link

and Extractive Module XT / Satellite XT.
Plug the cable (pictured green) from the
bottom plate of the Pyrolyzer Module
into the connector in the bottom of the
Extractive Module XT.
10.8

Power Connection

Use a standard power cable (obey any
local regulations) and bring it through the
power cable fitting into the module. Open
the module from the top and connect
Satellite XT 4-20 mA Versions
Pyrolyzer Module XT - Section 10
90 ...
240 VAC
10-4