Warning And Fault Messages; Warnings - Honeywell Satellite XT 9602-0200 4-20 mA Operation Manual

Satellite xt 4-20 ma series
Table of Contents

Advertisement

This section will help to determine the source of a fault or
7.1

Warning and Fault Messages

There are two general indications, warnings and faults, which require different attention.
If a fault or warning message does not appear in this section or if further assistance is
required, contact our service department.
7.1.1

Warnings

A warning condition indicates that the instrument requires some attention, but the
instrument is still able to monitor and operate as programmed.
The following will occur when the instrument detects a warning condition:
- the green status LED is flashing
- an alternating signal (2.8 to 4.0 mA, 0.1 Hz) is sent to external control systems;
depending on the configuration, this function can be enabled or disabled.
- depending on the configuration, the fault relay will be activated.
Press the <set> key to acknowledge and reset a Warning Condition.
L OW Z E R O
!
A s H 3
0 . 00
C H E C K S E N S O R !
A s H 3
0 . 00
Satellite XT 4-20 mA Versions
warning condition and will provide a corrective action.
This message indicates that the zero point of the sensor is out
of the optimum range.
To correct this problem, check the zero adjustment and
p p m
recalibrate, if necessary.
This message indicates that the sensor sensitivity does not
comply with the specifications.
The sensor must soon be replaced.
p p m
To receive this message, the function Auto Selftest must be
configured enabled; refer to Section 6, Configuration.
! This function is not available for instruments with
oxygen sensor or catalytic sensor.
Troubleshooting - Section 7
Check Sensor !
Low Zero !
7-1