Functional Test Md Models; Electrical Safety Test - Technogym Jog Now Excite+ 700 Technical Service Manual

Jog now excite+ series
Hide thumbs Also See for Jog Now Excite+ 700:
Table of Contents

Advertisement

JOG NOW Excite +: Service & maintenance manual - rev. 6.2
11.

FUNCTIONAL TEST MD MODELS

Every time maintenance and/or repairs are performed on a machine, it is necessary to carry out a
series of mechanical and electrical tests to ensure:
That the medical device is working properly;
That the medical device is working safely;
That the exercises are performed according to the biomedical specifications for which the
medical device was designed.
Technogym® ensures its own safety standards by means of the following tools and activities:
ISO 9001/2000 certification of the company;
Certification of the device in compliance with medical regulations 60601-1 and 60601-1-2;
Electrical testing of each device manufactured, according to the provisions of the
aforementioned standards;
Training of technical staff who perform technical service;
Use of checked and tested original Technogym® spare parts for all service operations;
CE Declaration of Conformity.

11.1. ELECTRICAL SAFETY TEST

Based on the risk analysis carried out by the After Sale department regarding installation and
technical service, it is still necessary to use a tester to perform a safety test on the grounding ring.
This test must ensure a continuous earth connection (resistance less than 4 4Ω (Ohm)) between the
grounding pole of the mains lead (connected to the machine) and the points listed below:
Pagina 11.1
Continued on following page...