Olympus FLUOVIEW FV1000 User Manual page 125

Laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

Image Acquisition (Spectral Image on XYL Image)
Image Acquisition (Spectral Image on XYL Image)
27
10
11
12
14
10. Set the range of wavelength to be
acquired, the slit width and the step.
Start = Start wavelength
End = End wavelength
Resolution = Slit width
StepSize = Step
11
Select AutoHV and then select
.
ScanSpeed.
*As the scan speed becomes slower,
noise can be removed while
maintaining the current brightness.
12. Select Lambda.
13.
Press the XYZ button to
acquire an image.
14. Click on SeriesDone, and "2D View-
LiveImage(x)" is displayed on the
window bar for the image that has
been acquired.