Olympus FLUOVIEW FV1000 User Manual page 124

Laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

Image Acquisition (Spectral Image on XYL Image)
Image Acquisition (Spectral Image on XYL Image)
4
The highest brightness wavelength
26
4. Click on the
5.
6. Press the XY Repeat button to start
5
7.
7
Note: Move the slit position only
8.
8
9.
button, and the
Spectral Setting window appears.
Set the slit width for CHS1 to 20
nm, for example.
scanning.
While observing the image, Click
the left side of slit
to the point which the highest
brightness is achieved.
while keeping the slit width
at 20 nm.
Adjust the image on the highest
brightness.
Press the Stop button to stop
scanning.
and drag
26