5 Getting Started

5 Getting Started

5.1
64

Turning on the PaX-i3D Green

To turn on the system, follow the steps below:
A. Before turning the equipment on, check whether the system is correctly
connected and installed (check the connection status between the equipment
and the PC).
B. Turn the ON / OFF switch underneath the handle frame to the ON position.
- If it has not been used for a long time, allow at least an hour
before the first exposure since turning on the equipment.
IMPORTANT
- Condensation could form inside the equipment if it is a
different temperature than the surrounding room. Only turn
on the equipment once it has reached room temperature.
- Wait at least 20 seconds after the equipment has been
turned off to turn it back on.
CAUTION
- Allow the equipment to warm up for at least 5 minutes
CAUTION
before acquiring images or preferably more than 30
minutes for image quality.
Do not position the patient in the unit while it is initiating.
The patient could be injured if the equipment malfunctions.
WARNING
C. Insert the license key for Ez3D plus
into the USB port on the back of PC.
The license key for Ez3D plus must
be attached to view or analyze the 3D
images.