4
Operation
4.7
Instrument controls
4.7
Instrument controls
Figure 4-5. Instrument control screenshow.
The Instrument Controls tab contains the controls for direct operation of the
instrument. At the top of the window, the user can Start the run, using whatever
parameters are currently present in the Experimental Design or Advanced
Experimental Design tabs.
Before clicking this button, it is wise to check that all parameters are correct and that a
valid, unique data file name has been entered. The software will double-check with the
user before allowing any files to be overwritten. The Stop button, which is available only
during a run, will abort the run immediately.
The Thermostat Control section allows for setting of the thermostat temperature,
which will be maintained during the MicroCal iTC
thermostatting the MicroCal iTC
shorter equilibration times. Also, high temperature thermostatting during cell cleaning
can improve the effects of the cleaning. Use the arrow buttons to raise or lower the
temperature or click in the text box and type in a new number. Click Set Jacket Temp to
set the temperature.
44
thermostatic (idle) state. Pre-
200
and samples at the run temperature will result in
200
MicroCal iTC
Operating Instructions 28-9639-77 AA