Maintenance
MicroCal iTC
Operating Instructions 28-9639-77 AA
200
WARNING!
Do not place containers of liquid on top of the instrument, except those
designed for the washing module. Spilled liquid is a fire and electrical
hazard.
CAUTION
Waste tubes and containers shall be secured and sealed to prevent
accidental spillage.
NOTICE
Never allow liquid in the cells to freeze. The expansion of the liquid can
distort the cells and rupture the most critical sensor, causing irreparable
damage.
NOTICE
Intermittent operation of the Washing Module is required so that the
Washing Module valves are on for a maximum of 20 minutes, then off for
30 minutes. Software limits the valve on period to 10 minutes before an
experiment, so this does not affect normal operation of the instrument.
NOTICE
The MicroCal iTC
instrument should always be moved in its normal
200
operating orientation. Other orientations will subject delicate sensors
inside the instrument to stress.
WARNING!
Replace fuses ONLY with same type fuses. Several spare fuses are
provided with the original shipment and the power receptacle is labeled
with the correct type.
WARNING!
Repairs, alterations or modifications must only be carried out by a GE
Healthcare specialist, or with explicit directions from a GE Healthcare
technician. Removal or modification of any cover or component could
result in an unsafe or easily damaged instrument. The GE Healthcare
service department will be happy to answer any questions and provide
parts and service when necessary.
WARNING!
Only spare parts that are approved or supplied by GE Healthcare may
be used for maintaining or servicing the system.
Safety instructions 2
Safety precautions 2.1
19