WARNING!
Power cord. Only use power cords delivered or approved by GE
Healthcare.
WARNING!
Do not block the ventilation inlets or outlets on the system.
WARNING!
The Washing Module may only be powered by the power supply
provided with the Unit.
WARNING!
Installing the controller. The controller should be installed and used
according to the instructions provided by the documentation included
in the shipment.
WARNING!
Replace fuses ONLY with 4.00 Amp 250 Volt Time Delay Fuses. Several
spare fuses are provided with the original shipment.
WARNING!
Access to power switch and power cord. Do not block the rear and side
panel of the instrument. The Power switch must always be easy to
access. The power cord must always be easy to disconnect.
NOTICE
Disconnect power. To prevent equipment damage, always disconnect
power from the MicroCal iTC
instrument module is removed or installed or a cable is connected or
disconnected.
WARNING!
All solutions in the cells must be cooled down below 40ºC before
removal. Any higher temperature may cause the syringe to break, and
will increase the dangers of most hazardous solutions.
and washing module system before an
200
MicroCal iTC
Operating Instructions 28-9639-77 AA