Appendix G: Shortening the cable

Pointek CLS200 Standard, Cable Version

: Possible only with the general purpose configuration; please verify against
CAUTION
product nameplate.
Preparation
Determine the required cable length, and subtract that amount from the actual length, to
find the excess length to cut off.
For example: 10 m (actual length) minus 9 m (required length) = 1 m (excess)
Steps
1.
Unscrew the cable gland compression nut to relieve the sealing cone and release
the cable.
7ML19985JH01
Pointek CLS200/300 Standard – INSTRUCTION MANUAL
actual
length
required
length
cable gland
compression nut
probe sleeve
excess
length
probe, lower
assembly
Page 65