Use with PD1000A Power Device Measurement System

Connecting the PD1000A Bias Ts to a Keysight B2902A SMU

Figure 5 below shows how to connect two PD1000A Bias T Networks to the
B2902A Source Measure Unit (SMU). Channel 1 on the SMU (front panel)
connects to the PD1000A Bias T for the RF input (Gate) to the device under test
(DUT). Channel 2 on the SMU (rear panel) connects to the PD1000A Bias T for
the RF output (drain or collector) of the DUT.
SMA to Port 1 on
Network Analyzer
B2902A CH1 Front Panel
to Gate Bias T
Figure 5
PD1000A Bias T Networks Connected to the B2902A SMU
Keysight PD1000A Bias T Operating Guide
B2902A CH2 Rear Panel to
Drain/Collector Bias T
PD1000A Bias T for
PD1000A DUT Test
the Gate
Fixture
SMA to Port 2 on
Network Analyzer
PD1000A Bias T for
Drain or Collector
17