Chapter 4 Explanation of Functions

■Interference Prevention

This function is used for preventing interference between 2 sensor heads.
Set this function when beam from the other sensor head enters into the mutual
interference area.
4
In case of connecting one sensor head, sampling cycle doubles when this
function is set to ON. Make sure to set this function to OFF when one sensor
head is connected.
● How to use
1
2
"8-6 Characteristics" – "Mutual Interference Area"
When connecting two sensor heads, there is the case that one sensor head
receives beam from the other sensor head, and this might cause a
measurement error if the setting distance between the two sensor heads is
close. In this case, by setting the Interference Prevention function to ON,
two sensor heads emit beam in turn and are not susceptible to the
interference of the other sensor head. The data update increases twice of the
setting sampling cycle.
Display "Interference Prevention" screen.
Switch the setting of interference
prevention.
The display is switched
OFF
each time the "Interference Prevention"
key is touched.
ON
and
4-58

Advertisement