Analog Output Parameter Settings - ABB FMT200-ECO2 Operating Instruction

Thermal mass flowmeter
Hide thumbs Also See for FMT200-ECO2:
Table of Contents

Advertisement

6.4 Analog output parameter settings

This page allows parameter settings of the analog output. For a better overview, the actual and the new
parameters are both shown.
The new settings will be sent to the FMT200-ECO2 (Sensyflow eco2) when the save-command is select-
ed from the „save data" menu.
After saving the data the FMT200-ECO2 (Sensyflow eco2) will initialize itself. During this save-initializa-
tion phase, the LED is blinking red and no measuring values will be sent.
(Please refer to chapter 6.3.7 "The configuration menu" on page 18 for further information.)
Fig. 6-16
– Analog output:
– Overflow 10%
– Range full scale value:
– Offset suppression:
– Filter factor:
– Output at error:
Signal
Min
Max
Up
Down
Values in brackets are valid when option „Overflow active" is chosen.
20
Thermal Mass Flowmeter FMT200-ECO2
Choose one output from the list.
With "Overflow active", a measuring value 10% higher than the adjusted "Range full scale value" can
be given out at the analog and frequency output.
Which flow rate shall correspond to the maximum output value.
Flow rates smaller than this limit will be displayed as zero flow. You should increase the flow suppres-
sion if you are operating with higher pressure.
damping of the analog output
How shall the FMT200-ECO2 (Sensyflow eco2) behave in case of an error?
Min:
output shows
Max:
output shows
Up:
output shows
Down:
output shows
0...20 mA
0 mA
20 (22) mA
> 22.5 mA
0%
100%
>100%
<
0%
4...20 mA
4 mA
20 (21.6) mA
> 22.5 mA
< 3.5 mA
Changing the configuration
0...5 V
0...10 V
0 V
0 V
5 (5.5) V
10 (11) V
> 5 V
42/14-41 EN

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sensyflow eco2

Table of Contents