GENERAL OPERATING INSTRUCTIONS
AccuPyc 1330
The following is a calibration report. The calibration standard used must be at least 10% of
nominal cell volume. We recommend 60 to 70% for the best results.
4-10
AccuPyc 1338 UX.XX
Serlal Hll .. ber XXXXX
CIlUbratlDn RepDrt
5.5775
ce
PaO'_ 1
Volume of Callbratlon Standarda
Nunber of purgesl 10
Equl11bratlon Rate.
0.0050
palg/.1n
Ce 11
Volulle
Deviation
Run
PaIr.
ce
ce
---------
--------
--------
1
12.3844
0.0008
2
12.3841
0.0004"
3
12.3841
0.0004
12.3634
-0.0003
5
12.3844
0.0007
6
12.3839
0.0002
7
12.3835
-0.0001
8
12.3831
-0.0006
9
12.3830
-0.0006
10
12.3828
-0.0009
Average Cell.Volumel
12.3837
ce
Average Expansion Volume.
7.9965
ce
Expansion
Volua.
ce
7.998e
7.9988
7.9993
7.9984
7.9991
7.9987
7.9984
7.9979
7.9980
7.9977
Standard Deviation.
Standard DeViation:
Figure
4-2.
Calibration Report
Devlat Ion
ce
0.0003
0.0003
0.0008
-0.0001
0.0006
0.0002
-0.0002
-0.0006
-0.0011'5
-0.0008
0.0006
cc
0.0005
ce
22
Aug 90