ACcuPyc 1330
and equation (9) rewritten as
VEXP
VSAMP
=
VCEll.- P
.:1& _ 1
Pz
g
APPENDIX 8
(12)
Equation (12) then becomes the working equation for the pycnometer. Calibration procedures are
provided to determine
V CEll.
and
V EXP
and the pressures are measured by a gauge pressure trans-
ducer. Provisions are made for conveniently charging and discharging gases at controlled rates,
for optimizing the relative sizes of the sample chambers and expansion volumes, and for cleans-
ing the samples of vapors which would render equations (1), (2), and (3) inadequate to describe
behavior.