Vega VEGABAR 51 Operating Instructions Manual page 38

Pressure transmitter, profibus pa
Hide thumbs Also See for VEGABAR 51:
Table of Contents

Advertisement

6 Set up with the indicating and adjustment module PLICSCOM
Carry out position cor-
rection
38
To select the adjustment unit (in the example switching over from bar
to mbar) you have to proceed as follows:
1
Push the [OK] button in the measured value display, the menu
overview is displayed.
Basic adjustment
Display
Diagnostics
Service
Info
2
Confirm the menu "Basic adjustment" with [OK], the menu item
"Unit" will be displayed.
Unit
Unit of measurement
bar
Temperature unit
°C
3
Activate the selection with [OK] and select "Units of measure-
ment with [->].
4
Activate the selection with [OK] and select the requested unit with
[->] (in the example mbar).
5
Confirm with [OK] and move to position correction with [->].
The adjustment unit is thus switched over from bar to mbar.
Proceed as follows to select the temperature unit:
à Activate the selection with [OK] and select "Temperature unit
with [->].
à Activate the selection with [OK] and select the requested unit with
[->] (e.g. °F).
à Confirm with [OK].
The temperature unit is hence switched over from °C to °F.
Proceed as follows:
1
Activate in the menu item "Position correction" the selection with
[OK].
Position correction
Offset
=
P
+0000 mbar
53 mbar
2
Select with [->], e.g. to accept actual measured value.
6)
Selection options: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, inH
mmH
O.
2
7)
Selection options: °C, °F.
6)
7)
VEGABAR 51 • Profibus PA
O,
2

Advertisement

Table of Contents
loading

Table of Contents