Preliminary Procedure - LeCroy ZS1000 Instruction Manual

High impedance active probe
Table of Contents

Advertisement

ZS1000 High Impedance Active Probe
Table 7-1. List of Required Equipment
Description
Digital Oscilloscope
ProBus interface; Windows-
based
Digital Multimeter (DMM)
4.5 digit
with test probe leads
DC: 0.1% Accuracy
AC: 0.1% accuracy
Function Generator
Sine Wave output amplitude
adjustable to 14.14 V
into 1 MΩ at 70 Hz
Power Supply
0-12 V, settable to 10 mV
Male to Male, 50 Ω, 36" Cable
BNC Coaxial Cable (2 ea.)
BNC Tee Connector
Male to Dual Female
Calibration Fixture
ProBus Extender Cable
50 Ω ± 0.05%
Terminator, Precision, BNC
Banana Plug Adapter (2 ea.) Female BNC to Dual Banana
Plug
BNC to Mini-grabber
BNC Male to Mini-grabber Cable,
36"

PRELIMINARY PROCEDURE

1. Connect the ZS1000 probe to the female end of the ProBus Exten-
sion Cable. Connect the male end of the ProBus Extension Cable to
channel 1 of the oscilloscope.
2. Turn the oscilloscope on and allow at least 30 minutes warm-up time
for the ZS1000 and test equipment before performing the Verifica-
tion Procedure.
3. Turn on the other test equipment and allow these to warm up for the
time recommended by the manufacturer.
4. While the instruments are reaching operating temperature, make a
photocopy of the Performance Verification Test Record (located in
Appendix A), and fill in the necessary data.
5. Select the channel to which the probe is connected. Set the oscillo-
scope scale factor to 20 mV/div.
7–2
Minimum Requirements
LeCroy WaveRunner Xi or
WaveSurfer Xs
Agilent Technologies 34401A or
Fluke 8842A-09
Agilent Technologies 33120A or
(5 V
)
Stanford Research Model DS340
p-p
rms
HP E3611A
Pomona 2249-C-36 or
Pomona 5697-36
Pomona 3285
LeCroy PROBUS-CF01
LeCroy TERM-CF01
Pomona 1269
Pomona 5187-C-36
Test Equipment Examples
ZS1000-OM-E Rev C

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zs1500

Table of Contents