Setup/Operation
Select Pipette Head Serial Number
Select the serial number for each installed pipette head from the drop-down lists, close
the rotating cover or hood door, and then select Next.
If the serial number is not displayed in the drop-down list, the pipette head must be
added:
1
Select Add new pipette head.
2
Do any of the following:
•
If using a tablet, select Scan and then select Camera. Optionally, select the icon
to switch between the front and rear camera. When prompted, scan the barcode
on the pipette head.
•
Enter the values from the Gilson Quality Control Report supplied with each
pipette head.
•
Using the optional barcode scanner (ordered separately), select Scan, and then
scan the barcode on the pipette head.
3
Select Save.
4
Repeat steps 2 and 3 for additional pipette heads. When finished adding heads,
select Back.
5
Select the serial number for each installed head, close the rotating cover or hood
door, and then select Next.
Gilson PIPETMAX® 268 User's Guide
2
2-19