Setup/Operation
Pipette Head and Tray Setup
Wait for the software to finish "Preparing pipette head" and then open the rotating cover or the hood door.
Remove Alignment Head (If Installed)
Remove the alignment head by disconnecting the terminal block connector from the rear
panel, and then removing the thumb nuts securing it to the upper pipette head
assembly. Cover the hole on the rear panel from the exterior using the black plug
supplied in the accessory package.
black plug
Connect External Safety Interlock Sensor
If your PIPETMAX does not have a rotating cover, connect the terminal block connectors for
the external safety interlock sensor to the input/output ports on the rear panel. They are
labeled TOP and BOTTOM to help ensure proper connection.
Install Tip Ejector Foot
If using a MAX8x20 or MAX8x200 pipette head, skip to
If using the MAX1000 pipette head, you ordered a tip ejector foot kit for eight and
one channel.
1
Using the supplied 3/32" hex wrench, remove the shoulder screw in the installed
tip ejector foot.
2
Remove the tip ejector foot.
3
Slide the tip ejector foot for eight and one channel onto the ejector leg.
4
Use the supplied 3/32" hex wrench to start threading the shoulder screw, but do not
fully tighten.
Gilson PIPETMAX® 268 User's Guide
Install Pipette Head(s)
on page 2-18.