Gilson PIPETMAX 268 User Manual page 19

Table of Contents

Advertisement

Introduction
Technical
Definition
Specification
Positioning
Machine Accuracy (without tips) in XYZ = +/-0.2 mm (0.008")
Performance
System Accuracy (with current tips) in XYZ = +/-0.9 mm (0.035")
Power
External Power Supply
Requirements
Voltage Input
Frequency: 50 to 60 Hz
Voltage: 100–240V AC
Voltage Output
Voltage: 24V DC
Current Rating: 6.25A, 150W
Removable Tray
9-position removable tray (microplate footprints, but not for 384-well microplates)
Capacity
9-position removable tray for 384-well microplates
Safety and
The PIPETMAX® 268 has been certified to safety standards specified for Canada, Europe, and the
Compliance
United States. Refer to the instrument rear panel label and the Declaration of Conformity document for
the current standards to which the instrument has been found compliant.
Volumetric
Specification for
Pipette Head
Transfers
MAX8x20
MAX8x200
Weight
24.9 kg (55 lbs.)
XYZ Motion
Parameters
Acceleration
Acceleration
Acceleration
Gilson PIPETMAX® 268 User's Guide
PIPETMAX® 268 (Continued)
Volume of
Distilled Water
1 μL
10 μL
20 μL
20 μL
100 μL
200 μL
X
Min
Power
1%
Speed
1 mm/sec
Speed Value
Y
Min
Power
1%
Speed
1 mm/sec
Speed Value
Z
Min
Power
1%
Speed
1 mm/sec
Speed Value
Maximum Permissible Errors
Systematic
Random
Error (μL)
Error (μL)
≤0.05
±0.08
≤0.10
±0.15
≤0.12
±0.25
≤0.16
±0.50
≤0.30
±1.00
≤0.50
±2.00
Max
100%
550 mm/sec
2
4000 mm/sec
Max
100%
550 mm/sec
4000 mm/sec
2
Max
100%
140 mm/sec
4000 mm/sec
2
PIPETMAX® 268 Technical Specifications (Page 2 of 2)
Systematic
Random
Error (%)
Error (%)
8
5
1.5
1
1.25
0.6
2.5
0.8
1
0.3
1
0.25
Default
50%
350 mm/sec
2
2200 mm/sec
Default
50%
350 mm/sec
900 mm/sec
2
Default
60%
120 mm/sec
650 mm/sec
2
1
1-5

Advertisement

Table of Contents
loading

Table of Contents