Setup/Operation
Step-by-step Wizard
1
Select
to go to the "Home" screen.
2
Select Run a protocol.
Select a Protocol
1
Select your protocol, and then select Next.
2
The Scanning dialog appears while the software checks that PIPETMAX is connected.
If connected, the software goes to the next screen. If not connected, a message will
appear suggesting possible solutions.
Variables
If your protocol has variables, enter a value for each, and then select Next.
When entering a value for a location, the following applies:
Use a colon (:) to specify a range.
Ranges can be numeric or alphanumeric.
Numeric ranges will be processed sequentially.
Alphanumeric ranges will be processed in a batch, if possible.
Use a comma (,) to separate values or ranges.
Specify the starting location first.
A dash (-) is not valid.
Labware Setup Guide
Select Step-by-step wizard.
The Step-by step wizard includes:
Gather Materials (Materials List)
Pipette Head and Tray Setup
Connect External Safety Interlock
Install Pipette
Head(s),
Select Pipette Head Serial
Review Initial Volumes
Run Protocol
Gather Materials (Materials List)
Gather the materials in the Materials list and a tip reload block for each rack, and then
select Next.
2-16
(which includes
Remove Alignment Head (If
Sensor,
Install Tip Ejector
Finish Tip Ejector Foot
Installation, and
Number)
Installed),
Foot,
Gilson PIPETMAX® 268 User's Guide