Weight
Receiver Input
Power Supply
Communication Range
Memory Storage
Re-Chargeable Battery Use
Charging Time
Operating Conditions/
Storage and Transport
Operating Altitude
Medium Priority Alarm Audible
Output
Moisture Protection
Limited Warranty
Safety Classification
No cleaning methods are recommended or tested for the receiver. The useful life of the receiver is 1 year. The
service/useful life for the accessories is 1 year. If you have difficulty reading your receiver in bright sunlight, you
may need to seek a shady location. Do not connect the receiver to any equipment not specified in IFU.
Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The receiver (MT22949) is intended for use in the electromagnetic environment specified in the next table. The
customer or the user of the receiver should ensure that it is used in such an environment.
Receiver Electromagnetic Immunity Specifications
IEC 60601 Test
Immunity Test
Level
Electrostatic Discharge
± 8 kV Contact
(ESD)
± 15 kV Air
IEC 61000-4-2
± 2 kV for power supply
Electrical Fast
Transient/Burst
± 1 kV for
IEC 61000-4-4
input/output lines
Surge
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
IEC 61000-4-5
Surge
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
IEC 61000-4-5
Voltage Dips, Short
U
0%
T
Interruptions and
U
0%
for 0.5 cycle at
Voltage Variations on
T
8 phase angles
Power Supply Input
Lines
U
70%
(30% dip in
T
25 cycles
IEC 61000-4-11
U
0%
for 250 cycles
T
IEC 60601-1-11
LBL014003 Rev 001 Dexcom G6 User Guide
4.0 ounces
5V DC, 1A
30 days of glucose data
7 days of tech support data
3 hours wall outlet
The device behaves normally while being charged
Do not hold the receiver while charging for over a minute
There are no risks to connecting any part of the system to an MSO (Multiple
Socket Outlet)
Temperature: 32° F-113° F
Humidity: 15%-95% RH (Storage/Transport 10%-95% RH)
-1200 feet to 13,500 feet
50 dB
IP22: Vertically falling drops
Protection against insertion of large objects and dripping water
Class II equipment
Transmitter
Compliance
Level
± 8 kV Contact
± 15 kV Air
± 2 kV for power supply
lines
lines
Not applicable
± 1 kV line(s) to line(s)
Not applicable
± 1 kV line(s) to line(s)
Not applicable
U
for 1 cycle
0%
for 1 cycle
T
U
0%
for 0.5 cycle at
T
8 phase angles
U
U
U
) for
70%
(30% dip in
T
T
25 cycles
U
0%
for 250 cycles
T
MT23681
20 feet
2 days
at 1 meter
SPL
1 year
Electromagnetic
Environment Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Main power quality should be that of a
typical commercial or hospital
environment.
Main power quality should be that of a
typical commercial or hospital
environment.
Main power quality should be that of a
typical commercial or hospital
environment.
Main power quality should be that of a
typical commercial or hospital
environment.
) for
T
21