Appendix G: Technical Information
G.2 Product Specifications
User is the single use operator in the home environment.
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Do not touch the metal connectors on the bottom of the transmitter and other open connectors on the receiver,
charging cable and charger.
Sensor Product Specifications
Glucose Range
Sensor Useful Life
Calibration
Calibration Range
Storage and Transport
Conditions
Sterilization
Transmitter Product Specifications
Part Number
Dimensions (Including
Sensor Pod)
Weight (Including Sensor
Pod)
Power Supply
Operational Conditions
Storage and Transport
Conditions
Operating Altitude
Limited Warranty/Useful
Life
IP28: Protection against insertion of large objects and immersion in water for up to
Moisture Protection
Protection Against
Electrical Shock
LBL014003 Rev 001 Dexcom G6 User Guide
40-400 mg/dL
Up to 10 days
Commercially available meter
20-600 mg/dL
Temperature: 36° F-86° F
Humidity: 10%-90% RH
Sterile by radiation
9445-02
Length: 1.8 inches
Width: 1.2 inches
Thickness: 0.6 inches
0.42 ounces
Silver oxide batteries (not replaceable)
Ambient temperature is 10° C-42° C (50° F-107.6° F)
Equilibrium temperature of less than 0.5° C (0.9° F) above ambient
Humidity: 10%-95% RH
Temperature: 32° F-113° F
Humidity: 10%-95% RH
-1300 feet to 13,800 feet
3 months
8 feet for 24 hours
Protection against submersion in water
Type BF applied part
19