Mounting Position; Mounting; Torques And Pressures - KROHNE OPTIFLUX 1000 Quick Start Manual

Electromagnetic flow sensor in sandwich design
Hide thumbs Also See for OPTIFLUX 1000:
Table of Contents

Advertisement

OPTIFLUX 1000

2.8.10 Mounting position

Figure 2-16: Mounting position

2.9 Mounting

2.9.1 Torques and pressures

The maximum pressure and torques values for the flowmeter are theoretical and calculated for
optimum conditions and use with carbon steel flanges.
Figure 2-17: Tightening of bolts
Tightening of bolts
• Always tighten the bolts uniformely and in diagonally opposite sequence.
• Do not exceed the maximum torque value.
• Step 1: Apply approx. 50% of max. torque given in table.
• Step 2: Apply approx. 80% of max. torque given in table.
• Step 3: Apply 100% of max. torque given in table.
01/2013 - 4002298501 - QS OPTIFLUX 1000 R04 en
www.krohne.com
INSTALLATION
2
13