KROHNE OPTIFLUX 6000 Quick Start Manual

KROHNE OPTIFLUX 6000 Quick Start Manual

Electromagnetic hygienic flow sensor, flanged and adapter versions for volumetric flowrate measurement of electrically conductive liquids
Hide thumbs Also See for OPTIFLUX 6000:

Advertisement

Quick Links

Quick Start
© KROHNE 03/2006
7.30954.31.00
JS
General
For in-depth information please consult
the handbook, data sheet, special
manuals and certificates supplied on
CD-ROM.
Installation, assembly, commissioning and
servicing must only be undertaken by
trained personnel. Maintenance which is
considered relevant to safety in the sense
of explosion protection must only be
carried out by the manufacturer, his
agents or under the supervision of
experts.
For use in hazardous areas, special
codes and regulations are applicable,
which are supplied in a separate
document that describes all hazardous
area relevant information.
Responsibility as to suitability and
intended use of this device rests solely
with the user. The supplier does not
accept any liability resulting from misuse
by the customer. Improper installation and
operation may lead to loss of warranty. In
addition, the "General conditions of sale",
found on the back of the invoice and
forming the basis of the purchasing
contract, are applicable.
If you need to return the device to the
manufacturer or supplier, please complete
the form given on the CD-ROM and attach
it to the device. KROHNE regrets that it
cannot repair or check your device unless
accompanied by this completed form.
OPTIFLUX
6000
Electromagnetic
Hygienic Flow Sensor
Flanged and adapter versions
for volumetric flowrate measurement
of electrically conductive liquids
Items supplied
Only separate version
Visual check
Liner materials
PFA
Teflon-PFA
Example
Electrode materials
HC4
Hastelloy C4
V4A
Stainless steel
Ti
Titanium

Advertisement

loading
Need help?

Need help?

Do you have a question about the OPTIFLUX 6000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for KROHNE OPTIFLUX 6000

  • Page 1 If you need to return the device to the manufacturer or supplier, please complete the form given on the CD-ROM and attach it to the device. KROHNE regrets that it cannot repair or check your device unless accompanied by this completed form.
  • Page 2: Handling And Installation Instructions

    OPTIFLUX 6000 Handling and installation instructions Quick Start OPTIFLUX 6000...
  • Page 3 149°F IFC 010 -25°C 60°C IFC 010 -13°F 140°F Note: The min. / max. temperatures for the compact devices are dependent on the liner / converter combination and the used type of adapter. See data sheet. Quick Start OPTIFLUX 6000...
  • Page 4: Special Considerations

    OPTIFLUX 6000 Special considerations Quick Start OPTIFLUX 6000...
  • Page 5 Metal pipelines, with internal coating, and internally non- without grounding rings. conductive pipelines. Grounding with grounding rings. Grounding rings Cleaning In principle, no special maintenance is needed. However, make sure that the used cleaning product doesn’t affect the outer surface and the gaskets. Quick Start OPTIFLUX 6000...
  • Page 6 2) Bolts to be used from material SS, class 70-A2. Max. Torque Step 1: approx. 50% of max. torque Step 2: approx. 80% of max. torque Step 3: 100% of max. torque given in tables above Quick Start OPTIFLUX 6000...
  • Page 7 OPTIFLUX 6000 Quick Start OPTIFLUX 6000...
  • Page 8 Croatia New Zealand KROHNE Messtechnik KROHNE Austria Ges.m.b.H. Singapore Denmark Pakistan GmbH & Co. KG Modecenterstraße 14 Tokyo Keiso - KROHNE Pte. Ltd. Ecuador Peru Ludwig-Krohne-Str. A-1030 Wien 27 Kian Teck Drive Jurong Egypt Poland D-47058 Duisburg TEL.: +43(0)1/203 45 32...