OPTIFLUX 1000

2.8 Installation conditions

2.8.1 Inlet and outlet

Use straight inlet and outlet pipe sections to prevent flow distortion or swirl, caused by bends
and T- sections
Figure 2-5: Recommended inlet and outlet section
1 Refer to chapter "Bends in 2 or 3 dimensions"
2 ≥ 2 DN

2.8.2 Bends in 2 or 3 dimensions

Figure 2-6: 2 and 3 dimensional bends, in front of flowmeter
1 Bends in 2 dimensions: ≥ 5 DN; bends in 3 dimensions: ≥ 10 DN

2.8.3 T-section

Figure 2-7: Distance behind a T-section
1 ≥ 10 DN