Testing Time Interval Accuracy

Testing the time interval accuracy does not check a specification, but
does verify proper operation of the 125-MHz oscillator.
This test verifies that the 125-MHz timing acquisition synchronizing
oscillator is operating within limits.
Table 15
Equipment Required
Equipment
Time Mark Generator
Oscilloscope
BNC Tee
BNC Cable (Qty 2)
BNC Test Connector,
1x2
Critical Specifications
Rise time < 10 ns; device
under test error is indicated
to within one least
significant digit
100 MHz bandwidth
Recommended
Model/Part
Tektronix TG 501A or
equivalent
HP 54600
HP 1250-0781
HP 10503A
173