HP 54620A User's And Service Manual page 202

Logic analyzer
Hide thumbs Also See for 54620A:
Table of Contents

Advertisement

External
Input R & C
trigger
Maximum Input
Trigger Threshold
Threshold Accuracy
Minimum Input Change
Minimum Pulse Width
Trigger
Output is a rising edge at the trigger point.
Output
Output Level
Delay, data in to
trigger out
Jitter
Maximum Output Rate
1
Performance specification, tested by threshold test. See Chapter 5, "Testing, Adjusting, and Troubleshooting
the Analyzer."
Performance Characteristics
~ 1 Meg Ω and 12 pF, compatible with HP 1007X
probes.
± 40 V peak.
± 6 volts, settable in 50-mV increments.
1
± 100 mV or 6% of setting, whichever is greater.
200 mV pp.
20 ns.
0 to ≥ 2.0V into 50 Ω
0 to ≥ 4.8V open circuit.
~ 85 ns.
± Sample period + 10 ppm.
≅ 2 kHz (HP 54620A), ≅ 1.5 kHz (HP 54620C) with the
analyzer stopped, 20/sec running.
201

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

54620c

Table of Contents