Test Procedures; Electrical Testing; Tools Required - Ransburg 19372 Service Manual

No.2 process handgun electric motor version
Hide thumbs Also See for 19372:
Table of Contents

Advertisement

TEST PROCEDURES

Fluid Testing
Needle Locknut Adjustment
When the paint does not shut-off after the trigger
has been released, adjust the knurled nuts (in
front of trigger). If this does not correct the prob-
lem, refer to the "Troubleshooting Guide" in the
"Maintenance" section.
1. Turn the outer locknut a few times count-
er-clockwise. DO NOT thread it off the needle.
2. Adjust the inner locknut counter-clockwise, to
allow the needle to completely seat, and hold
the trigger firmly.
3. Make further adjustments of the outer and inner
locknuts as needed for fine adjustment of fluid
delivery with thin viscosity paints.
† Primary fluid pressure control is from
the paint pot regulator.

ELECTRICAL TESTING

If an electrical malfunction should occur in the
system, check the hand gun, high voltage cable,
supply fuses, and the power source according to
the prescribed test procedures for each. After
these electrical tests have been completed and it
is determined that a fault may exist in the power
supply, make the high voltage short circuit cur-
rent test as indicated in the power supply service
manual.
† These tests should be made as ad-
vised under the "Maintenance" section of
the manual.
AH-13-01.3
No. 2 Process Electric Motor Version - MAINTENANCE
NOTE
NOTE

TOOLS REQUIRED

76652-02 or 76652-04 Ransburg
Testers Kit
Gun Voltage Output (SCI)
The Short Circuit Current (SCI) will determine if
the high voltage portion of the system is operat-
ing correctly by shorting the maximum current
flow to ground. The SCI reading on either the
power supply current meter and/or the externally
connected meter at the contact brush should
be between 80 and 115 microamperes (.08 and
.115 milliamperes). Readings outside this range
indicate possible component failure and indicate
the need for further testing and/or repair. There
are two procedures that can be used to perform
an SCI Test..
NOTE
† If a proper reading is not obtained, per-
form the same test on a spare gun (if one
is available). If the spare gun test reading
is normal, refer to the "Troubleshooting
Guide" in the "Maintenance" section. If no
spare gun is available, perform a short cir-
cuit cur-rent test of the power supply.
† When the motor circuit is activated the
shaft will be rotating and the HIGH VOLT-
AGE WILL BE ON.
External Meter Test
1. This method uses the test assembly (76652-
02 or 76652-04). It must be used if the power
supply DOES NOT have a microampmeter.
2. Remove the bell from the gun. Connect one
of the alligator clamps of the test meter to the
contact brush and the other to ground.
3. Set the switch on the test meter to µA
(microamperes).
4. Turn the power supply ON.
5. Activate the motor circuit at the gun. The test
meter reading should be approximately 80 to
115 microamperes.
24

Advertisement

Table of Contents
loading

Table of Contents