S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual

173 CB_MULTITIMER

Message: System supports only four timer(TIMER1, ..., TIMER4).
Remarks: The LPT function was expecting TIMER1, TIMER2, TIMER3, or TIMER4 for the instr_id
parameter, but some other instrument ID was sent (for example, TIMER5 or SMU1).
194 MX_INVLDTRM
Message: Invalid terminal: XXX.
Remarks: The terminal specified is invalid for the command.
233 FM_NOCON
Message: Cannot force when not connected.
Remarks: The instrument must be connected to a device under test (DUT) before it can be used.
455 ECP_PROTOVER
Message: Protocol version mismatch.
Remarks: The application controller software version and the tester software version do not match.
Either the system software was not installed correctly, or the application controller is being used to
control an older or newer tester not intended to be used with this application controller.
601 SYS_INTERNAL_ERR
Message: System software internal error.
Remarks: An internal system software error has occurred.
610 SYS_SPAWN_ERR
Message: Could not start XXX.
Remarks: An internal system software error has occurred.
611 SYS_NETWORK_ERR
Message: Network error.
Remarks: The system is having problems communicating over the network. Make sure that all
network connections are secure, all network cables are intact, and any network routers are
functioning properly.
612 SYS_PROTOCOL_ERR
Message: Protocol error.
Remarks: An internal system software error has occurred.
S530-900-01 Rev. E / September 2017
Section 4: Error definitions
4-5