S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual
Also see
rtfary
(on page 3-63)
savgX
(on page 3-64)
sintgX
(on page 3-82)
smeasX
(on page 3-84)
trigXg, trigXl
This command monitors for a predetermined level of voltage, current, conductance, capacitance, or time.
Usage
int trigig(int instr_id, double value);
int trigil(int instr_id, double value);
int trigtg(int instr_id, double value);
int trigtl(int instr_id, double value);
int trigvg(int instr_id, double value);
int trigvl(int instr_id, double value);
instr_id
value
Details
The trigXl and trigXg commands are used with the searchX command or with one of the sweep
measurement commands: smeasX, sintgX, or savgX.
•
The trigXg or trigXl command provides the sweepX command the digital feedback to allow
for the increase or decrease in sourcing values.
•
The trigXl and trigXg commands must be located before any associated searchX
commands.
•
Triggers are not automatically reset by the searchX or sweepX command. A single call to the
trigXl or trigXg command can be followed by two or more calls to the searchX or sweepX
commands.
The specified trigger point is automatically cleared when a clrtrg or devint command is executed.
S530-900-01 Rev. E / September 2017
The instrument identification code of the monitoring instrument
The voltage, current, conductance, capacitance, or time specified as the trigger
point; this trigger point value is reached when either of the following occurs:
•
The measured value is equal to or greater than the value argument of the
trigXg command
•
The measured value is less than the value argument of the trigXl
command
Section 3: LPTLib command reference