Section 4: Error definitions

129 UT_TMRIVLD

Message: Timer not ENABLED.
Remarks: Time measurements can only be made on a timer when the timer is enabled.
137 UT_INVLDVAL
Message: Invalid value for modifier.
Remarks: An invalid option was used for the setmode or getstatus command.
152 CB_BADFUNC
Message: Function not supported by XXX (NNN).
Remarks: The driver for this instrument does not support the Linear Parametric Test Library (LPTLib)
command used.
156 CB_NOFILE
Message: Configuration file does not exist: XXX.
Remarks: A required configuration file is missing.
157 CB_FORMAT
Message: Configuration file format error. File: XXX, Section: XXX, Key: XXX.
Remarks: A system configuration file has a missing entry or an entry that was formatted in an
unexpected way.
162 CB_INVLDERROR
Message: Invalid error number: NNN.
Remarks: The logging .ini file has a format error.
163 CB_INVLDEVENT
Message: Invalid event number: NNN.
Remarks: The logging.ini file has a format error.
166 CB_INSNOTREC
Message: Instrument with model code XXX is not recognized.
Remarks: An instrument is not recognized by the system. The instruments may be misread over the
170 CB_INITFAIL.
4-4
S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual