Keithley S530 Manual page 102

Parametric test system, kte linear parametric test library
Hide thumbs Also See for S530:
Table of Contents

Advertisement

Section 3: LPTLib command reference
Modifiers that affect CVU measurements
Modifier
KI_CVU_ACV
KI_CVU_ACZ_RANGE
KI_CVU_CORRECT LOAD
KI_CVU_CORRECT_OPEN
KI_CVU_CORRECT_SHORT
KI_CVU_FREQ
KI_CVU_LENGTH
KI_CVU_MODE
KI_CVU_MODEL
KI_CVU_SPEED
Modifiers that affect matrix operation
Modifier
KI_AUTO_GND_PINS
3-80
S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual
Value
<value>
<value>
0 or 1
0 or 1
0 or 1
<value>
KI_CVU_CABLE_CORR_LV
KI_CVU_CABLE_CORR_HV
KI_CVU_CABLE_CORR_LV_HV
0 or 1
0 to 5
0 to 2
Value
70xB: 0 or 1
Comment
10 mV to 100 mV (45 mV is default)
0 = Autorange (default)
1e-6 = 1 µA range
30e-6 = 30 µA range
1e-3 = 1 mA range
0 = Off (default)
1 = On
0 = Off (default)
1 = On
0 = Off (default)
1 = On
1 kHz to 2 MHz (100 kHz is default)
Specifies the type of cables used in
the system so that cable-length
corrections can be automated.
0 = User mode
1 = System mode (default)
0 = Z, theta
1 = R, jx
2 = Cp, Gp (default)
3 = Cs, Rs
4 = Cp, D
5 = Cs, D
0 = Fast (default)
1 = Normal
2 = Quiet
Does not affect the intgc, intgg,
or intgcg LPT commands. To
adjust the speed of these LPT
commands, use the KI_INTGPLC
setmode.
Comment
Enable or disable the feature that
automatically connects all unused
pins to GND (including the
CHUCK).
0 = Disable
1 = Enable
S530-900-01 Rev. E / September 2017

Advertisement

Table of Contents
loading

This manual is also suitable for:

S540

Table of Contents