Calibration Overview - Renishaw ATOM Installation Manual

Rotary encoder systems
Hide thumbs Also See for ATOM:
Table of Contents

Advertisement

Calibration overview

Calibration is an essential operation that completes readhead set-up, with the optimum incremental and reference mark signal settings stored in the readhead's non-volatile memory.
This section is an overview of the calibration procedure for an ATOM system.
More detailed information on calibrating the system is on page 17.
Ensure AGC is switched off.
Systems are shipped with AGC switched off.
When AGC is switched off the readhead set-up LED will be Green, Orange or Red.
Align the readhead to maximise signal strength over the full rotation.
The readhead set-up LED will be Green.
Initiate Incremental signal calibration routine.
The readhead set-up LED will be single flashing Blue.
Rotate the axis at slow speed (<100 mm/s).
The readhead set-up LED will start double flashing Blue.
Reference mark not used
Reference mark
Manually exit calibration routine.
Rotate the readhead back and forth over the reference mark
The readhead set-up LED will stop flashing Blue.
until the readhead set-up LED stops flashing Blue.
The system is now calibrated and ready for use.
AGC can now be switched on if required.
NOTE: If calibration fails (readhead set-up LED continues flashing Blue),
restore factory defaults, check the readhead installation and repeat the calibration routine.
ATOM
16
rotary installation guide