Monitoring Uterine Activity Internally; Parts Required; Directions For Use Of Iupc - EDAN F6 User Manual

F series festal & material monitor
Hide thumbs Also See for F6:
Table of Contents

Advertisement

F Series Fetal & Maternal Monitor User Manual
4) Check that the value on the display shows this change in pressure.
If a TOCO transducer fails the test, repeat this test with another transducer. If the second one
passes the test, defect of the first transducer is confirmed. Replace it with a good one. If the
second transducer fails the test as well, contact the manufacturer for service.

7.6 Monitoring Uterine Activity Internally

7.6.1 Parts Required

1) Disposable intrauterine pressure catheter ACCU-TRACE
2) Reusable intrauterine pressure cable ("IUP cable" for short)
3) Reusable intrauterine pressure connecting cable ("connecting cable" for short)
The following illustration shows how these parts should be connected:
(1)

7.6.2 Directions for Use of IUPC

Preparation
1)
Gather supplies: ACCU-TRACE IUPC, reusable cable, and amnioinfusion supplies if
needed.
2)
Open the sterile ACCU-TRACE IUPC package.
Figure 7-6 Testing a TOCO Transducer
(2)
Figure 7-7 Connection for IUP Monitoring
- 65 -
Fetal Monitoring
IUPC ("IUPC" for short)
(3)

Advertisement

Table of Contents
loading

This manual is also suitable for:

F9 expressF9

Table of Contents