A7.4 Recommended Separation Distances

portable and mobile RF communications equipment and the
The F Series Fetal & maternal Monitor is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the F Series Fetal & maternal
Monitor can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the F Series Fetal & maternal
Monitor as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Recommended separation distances between
F Series Fetal & maternal Monitor
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
- 179 -
EMC Information
800 MHz to 2.5 GHz
d
2
3 .
P
0.23
0.73
2.3
7.3
23