9.2.5 SI (Signal Intensity)*

*Only applicable to the EDAN module
The signal intensity (SI) shows perfusion in numeric, and it reflects the pulse intensity of the
measurement site. The SI ranges from 0 to 10, with a larger value indicating the more intense
signal. When the SI value reaches 10, the signal quality is optimal. If the SI value is less than 2, it
indicates that the pulse at the current site is weak, and you should change the measurement site.
The SI value is displayed in the SpO
9.2.6 Switching the SpO
You can choose to switch the SpO
1 Select the setup key
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select SpO
.
2
4 Select ON (default) or OFF for Alarm.
5 Select OK.
9.2.7 Changing SpO
You can change the SpO
1 Select the setup key
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select SpO
.
2
4 Select a value from 50 ~ 99 for Low Alarm Limit. (The step is 1, and the default value is
90%.)
5 Select a value from 51 ~ 100 for High Alarm Limit. (The step is 1, and the default value is
100%.)
6 Select OK.
parameter area.
2
Alarm On or Off
2
alarm on or off.
2
on the main interface.
Alarm Limits
2
alarm limits.
2
on the main interface.
- 93 -