9.3 Maternal HR Monitoring

9.3.1 Introduction

Maternal heart rate (HR) monitoring does not need an extra accessory. When you perform ECG
or SpO
(Pulse) monitoring, the HR result can be acquired at the same time.
2
When monitoring ECG and SpO
them is being performed, the source will automatically switch to the available one (the screen
reading should prevail).

9.3.2 Choosing HR Source

You can change the HR source.
1 Select the setup key
2 Select Mother > HR Source.
3 Select ECG (default) or Pulse (during SpO
4 Select OK.

9.3.3 Changing HR Beep Volume

When the HR beep is enabled, the monitor gives a beep sound of maternal heart.
To change the HR beep volume,
1 Select the setup key
2 Select Mother > HR Beep.
3 Select OFF (default), Low or High.
4 Select OK.
NOTE:
1 The DECG beep and HR beep share the same audio channel. Once the HR beep is
switched on, the DECG beep is disabled (set to OFF) automatically.
2 Once the DECG/HR beep volume is changed, the sound switches to channel 1
automatically. Therefore, it is advised against changing DECG/HR beep volume in the
monitoring process.
at the same time, you can choose the HR source. If only one of
2
on the main interface.
monitoring).
2
on the main interface.
- 94 -
Maternal Monitoring