A2.4 Recommended Separation; Distances - EDAN H10 User Manual

Finger oximeter
Table of Contents

Advertisement

H10 Finger Oximeter User Manual
A2.4

Distances

Recommended separation distances between
portable and mobile RF communications equipment and H10
H10 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of H10
can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications
equipment (transmitters) and H10 as recommended below, according to
the maximum output power of the communications equipment.
Rated
maximum
output
power
of
transmitter(W)
0.01
0.1
1
10
100
Recommended
Separation
distance
transmitter(m)
80 MHz to 800 MHz
d
/
/
/
/
/
/
according
P
1
.
2
0.12
0.38
1.2
3.8
12
- 76 -
Separation
to
frequency
800 MHz to 2.7 GHz
d
P
2
3 .
0.23
0.73
2.3
7.3
23
of